Central
Remote Recording Control
Customers can use the Remote Recording Control feature in Central's File Storage application to remotely begin, end, and pause their recordings. This feature utilizes the Neural Signal Processor's (NSP) or Cereplex Direct's digital input port on the ...
NEV and NSx File Specifications
See the attached PDF for the current file specifications. Below you'll find a screenshot of our short summaries for the currently available file specifications in Central 7.0.6:
Packet Loss Error Message
Introduction When using the Central Software, it is possible to receive a message that indiciates packet drop or packet loss. This indicates that some data was lost between the data acquisition system and the PC. Since this error typically is due to ...
Multiple Instances of Central on the Same Host PC
NOTICE: This feature has been removed as of Central version 7.5.0. 7.5.0 and later supports running two NSPs in one instance of Central, but cannot run 3 or more in one instance. If you prefer having separate instances for each NSP connected to your ...
Recording Restarted Error
CauseIf you're experiencing a warning dialogue with the message "File Recording Restarted" then it means that HDDs inside the host PC can not keep up with the amount of data they are receiving from the NSP. Specifically, it indicates that the buffer ...